Proline Promass 83E
Coriolis flowmeter

Coriolis flowmeter - Proline Promass 83E ©Endress+Hauser

The flowmeter for minimized cost of ownership with extended transmitter functionality

Promass E has a long standing reputation as a cost efficient solution for Coriolis applications. Combined with the Promass 83 transmitter with touch control, four line display and extended functionality like software options for filling and dosing, concentration measurement or advanced diagnostics, Promass 83E offers highly accurate measurement of liquids and gases for a wide range of applications.

Dette produktet er ikke tilgjengelig for dette markedet

  • Fordeler

    • Cost-effective – multi-purpose device; an alternative to conventional volumetric flowmeters

    • Fewer process measuring points – multivariable measurement (flow, density, temperature)

    • Space-saving installation – no in/outlet run needs

    • Quality – software for...

  • Bruksområde

    • Measuring principle operates independently of physical fluid properties such as viscosity or density

    • Highly accurate measurement of liquids and gases for a wide range of standard applications.

    Device properties

    • Compact dual-tube system

    • Medium...

Funksjoner og spesifikasjoner

Tilbake

Details

  • Measuring principle

    Coriolis

  • Product headline

    The flowmeter for minimized cost of ownership with extended transmitter functionality. Highly accurate measurement of liquids and gases for a wide range of standard applications.

  • Sensor features

    Cost‐effective – multi‐purpose device; an alternative to conventional volumetric flowmeters. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in/outlet run needs. Compact dual‐tube system. Medium temperature up to +140 °C (+284 °F).

  • Transmitter features

    Quality – software for filling & dosing, density & concentration, advanced diagnostics. Flexible data transfer options – numerous communication types. Automatic recovery of data for servicing. 4‐line backlit display with touch control. Device in compact or remote version.

  • Nominal diameter range

    DN 8 to 80 (⅜ to 3")

  • Wetted materials

    Measuring tube: 1.4539 (904L)
    Connection: 1.4404 (316/316L)

  • Measured variables

    Mass flow, density, temperature, volume flow, corrected volume flow, reference density

  • Max. measurement error

    Mass flow (liquid): ±0.15 % (standard), 0.10 % (option)
    Volume flow (liquid): ±0.15 %
    Mass flow (gas): ±0.75 %
    Density (liquid): ±0.0005 g/cm³

  • Measuring range

    0 to 180 000 kg/h (0 to 6600 lb/min)

  • Max. process pressure

    PN 100, Class 600, 63K

  • Medium temperature range

    –40 to +140 °C (–40 to +284 °F)

  • Ambient temperature range

    Standard: –20 to +60 °C (–4 to +140 °F)
    Option: –40 to +60 °C (–40 to +140 °F)

  • Sensor housing material

    1.4301 (304), corrosion resistant

  • Transmitter housing material

    Powder‐coated die‐cast aluminium
    1.4301 (304), sheet
    CF3M (316L), cast

  • Degree of protection

    IP67, type 4X enclosure. Remote transmitter: IP67, type 4X enclosure

  • Display/Operation

    4‐line backlit display with touch control (operation from outside)
    Configuration via local display and operating tools possible

  • Outputs

    4 modular outputs:
    0‐20 mA (active)/4‐20 mA (active/passive)
    Pulse/frequency/switch output (passive)
    Relay

  • Inputs

    2 modular inputs:
    Status
    0‐20 mA (active)/4‐20 mA (active/passive)

  • Digital communication

    HART, PROFIBUS PA/DP, FOUNDATION Fieldbus, Modbus RS485, EtherNet/IP

  • Power supply

    DC 16 to 62 V
    AC 85 to 260 V (45 to 65 Hz)
    AC 20 to 55 V (45 to 65 Hz)

  • Hazardous area approvals

    ATEX, IECEx, FM, CSA, NEPSI, TIIS

  • Other approvals and certificates

    3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR, SIL, marine
    PED, CRN
    3-A

  • Product safety

    CE, C-tick, EAC marking

  • Functional safety

    Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511

  • Metrological approvals and certificates

    Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR

  • Marine approvals and certificates

    Marine approval

  • Pressure approvals and certificates

    PED, CRN

  • Material certificates

    3.1 material

  • Measuring principle

    Coriolis

  • Characteristic / Application

    An alternative to conventional volumetric flowmeters, this flowmeter has a low "cost of ownership".

  • Ambient temperature

    -20...+65°C
    (-4...+140°F)

  • Process temperature

    -40...+125°C
    (-40...+257°F)

  • Process pressure absolute

    PN 40...100
    Cl150...600
    JIS 10...63K

  • Wetted parts

    904L/1.4539

  • Output

    4...20mA
    Pulse/Frequency (10KHz, active/passive)
    Relays/Status

  • Certificates / Approvals

    ATEX
    FM
    CSA

  • Measuring principle

    Coriolis

  • Product headline

    The flowmeter for minimized cost of ownership with extended transmitter functionality. Highly accurate measurement of liquids and gases for a wide range of standard applications.

  • Sensor features

    Cost‐effective – multi‐purpose device; an alternative to conventional volumetric flowmeters. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in/outlet run needs. Compact dual‐tube system. Medium temperature up to +140 °C (+284 °F).

  • Transmitter features

    Quality – software for filling & dosing, density & concentration, advanced diagnostics. Flexible data transfer options – numerous communication types. Automatic recovery of data for servicing. 4‐line backlit display with touch control. Device in compact or remote version.

  • Nominal diameter range

    DN 8 to 80 (⅜ to 3")

  • Wetted materials

    Measuring tube: 1.4539 (904L)
    Connection: 1.4404 (316/316L)

  • Measured variables

    Mass flow, density, temperature, volume flow, corrected volume flow, reference density

  • Max. measurement error

    Mass flow (liquid): ±0.15 % (standard), 0.10 % (option)
    Volume flow (liquid): ±0.15 %
    Mass flow (gas): ±0.75 %
    Density (liquid): ±0.0005 g/cm³

  • Measuring range

    0 to 180 000 kg/h (0 to 6600 lb/min)

  • Max. process pressure

    PN 100, Class 600, 63K

  • Medium temperature range

    –40 to +140 °C (–40 to +284 °F)

  • Ambient temperature range

    Standard: –20 to +60 °C (–4 to +140 °F)
    Option: –40 to +60 °C (–40 to +140 °F)

  • Sensor housing material

    1.4301 (304), corrosion resistant

  • Transmitter housing material

    Powder‐coated die‐cast aluminium
    1.4301 (304), sheet
    CF3M (316L), cast

  • Degree of protection

    IP67, type 4X enclosure. Remote transmitter: IP67, type 4X enclosure

  • Display/Operation

    4‐line backlit display with touch control (operation from outside)
    Configuration via local display and operating tools possible

  • Outputs

    4 modular outputs:
    0‐20 mA (active)/4‐20 mA (active/passive)
    Pulse/frequency/switch output (passive)
    Relay

  • Inputs

    2 modular inputs:
    Status
    0‐20 mA (active)/4‐20 mA (active/passive)

  • Digital communication

    HART, PROFIBUS PA/DP, FOUNDATION Fieldbus, Modbus RS485, EtherNet/IP

  • Power supply

    DC 16 to 62 V
    AC 85 to 260 V (45 to 65 Hz)
    AC 20 to 55 V (45 to 65 Hz)

  • Hazardous area approvals

    ATEX, IECEx, FM, CSA, NEPSI, TIIS

  • Other approvals and certificates

    3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR, SIL, marine
    PED, CRN
    3-A

  • Product safety

    CE, C-tick, EAC marking

  • Functional safety

    Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511

  • Metrological approvals and certificates

    Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR

  • Marine approvals and certificates

    Marine approval

  • Pressure approvals and certificates

    PED, CRN

  • Material certificates

    3.1 material

  • Hygienic approvals and certificates

    3-A

  • Measuring principle

    Coriolis

  • Product headline

    The flowmeter for minimized cost of ownership with extended transmitter functionality. Highly accurate measurement of liquids and gases for a wide range of standard applications.

  • Sensor features

    Cost‐effective – multi‐purpose device; an alternative to conventional volumetric flowmeters. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in/outlet run needs. Compact dual‐tube system. Medium temperature up to +140 °C (+284 °F).

  • Transmitter features

    Quality – software for filling & dosing, density & concentration, advanced diagnostics. Flexible data transfer options – numerous communication types. Automatic recovery of data for servicing. 4‐line backlit display with touch control. Device in compact or remote version.

  • Nominal diameter range

    DN 8 to 80 (⅜ to 3")

  • Wetted materials

    Measuring tube: 1.4539 (904L)
    Connection: 1.4404 (316/316L)

  • Measured variables

    Mass flow, density, temperature, volume flow, corrected volume flow, reference density

  • Max. measurement error

    Mass flow (liquid): ±0.15 % (standard), 0.10 % (option)
    Volume flow (liquid): ±0.15 %
    Mass flow (gas): ±0.75 %
    Density (liquid): ±0.0005 g/cm³

  • Measuring range

    0 to 180 000 kg/h (0 to 6600 lb/min)

  • Max. process pressure

    PN 100, Class 600, 63K

  • Medium temperature range

    –40 to +140 °C (–40 to +284 °F)

  • Ambient temperature range

    Standard: –20 to +60 °C (–4 to +140 °F)
    Option: –40 to +60 °C (–40 to +140 °F)

  • Sensor housing material

    1.4301 (304), corrosion resistant

  • Transmitter housing material

    Powder‐coated die‐cast aluminium
    1.4301 (304), sheet
    CF3M (316L), cast

  • Degree of protection

    IP67, type 4X enclosure. Remote transmitter: IP67, type 4X enclosure

  • Display/Operation

    4‐line backlit display with touch control (operation from outside)
    Configuration via local display and operating tools possible

  • Outputs

    4 modular outputs:
    0‐20 mA (active)/4‐20 mA (active/passive)
    Pulse/frequency/switch output (passive)
    Relay

  • Inputs

    2 modular inputs:
    Status
    0‐20 mA (active)/4‐20 mA (active/passive)

  • Digital communication

    HART, PROFIBUS PA/DP, FOUNDATION Fieldbus, Modbus RS485, EtherNet/IP

  • Power supply

    DC 16 to 62 V
    AC 85 to 260 V (45 to 65 Hz)
    AC 20 to 55 V (45 to 65 Hz)

  • Hazardous area approvals

    ATEX, IECEx, FM, CSA, NEPSI, TIIS

  • Product safety

    CE, C-tick, EAC marking

  • Functional safety

    Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511

  • Metrological approvals and certificates

    Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR

  • Marine approvals and certificates

    Marine approval

  • Pressure approvals and certificates

    PED, CRN

  • Material certificates

    3.1 material

Dokumenter / håndbøker / programvare

Tilbehør / reservedeler

Nr
Order code
Specification
  • Erstatter