Du må skrive inn minst 2 tegn for å starte søket.

Analog pH sensor
Ceraliquid CPS41

Ceraliquid CPS41 - Analog pH sensor with refillable liquid KCl electrolyte ©Endress+Hauser

The pH-value of a liquid can be calculated using the potentiometric measurement principle. This video shows what it is about and how this measuring principle works.

Glass electrode for applications with fast-changing medium compositions or low conductivity

Ceraliquid CPS41 is the analog high performer for pH measurement in harsh chemical applications and media with low conductivity or a considerable content of organic solvents. The sensor is designed for fast response guaranteeing you high process safety even in applications with quickly changing medium compositions.

Hurtigvalg

Velg ett av våre foreslåtte alternative
  • <%label%>
  • <%feature%>
  • Process Connection
  • <%message%>

eller

individuell konfigurasjon

Fortsett for å velge blant alle mulige alternativer
  • Fordeler

    • Resistant to poisoning due to constant refilling of KCl bridge electrolyte and separate reference lead

    • Applicable at very low conductivities (= 0.1 µS/cm) thanks to liquid KCl electrolyte

    • Suitable for cleaning in place (CIP) and sterilization in place (SIP)

    • Perfectly suited for quickly changing media: Combination of liquid...

  • Bruksområde

    Media with very low conductivities or a considerable amount of organic solvents or alcohols and fast changing media:

    • Chemical industry

    • Power plants, e.g. boiler feed water

    • Life sciences, e.g. water for injection

    With ATEX, FM and CSA approval for use in...

Funksjoner og spesifikasjoner

Tilbake

Details

  • Measuring principle

    Glass electrode

  • Application

    - special applications with high requirements with regard to accuracy and speed - rapid changing composition of media, highly clogging media, low conductivities.

  • Characteristic

    - for lowest conductivity - diaphragm with defined electrolyte flow - can be subject to pressure to prevent clogging - gel-free refillable electrolyte - fast response time

  • Measurement range

    pH 0-14

  • Measuring principle

    liquid-KCl with ceramic diaphragm

  • Design

    - all shaft lengths with temperature sensor.

  • Material

    glass / ceramic

  • Dimension

    Diameter: 12mm (0.46 inch) shaft length: 120, 225, 360 and 425 mm
    (4.68, 8.77, 14.04 and 16,57 inch)

  • Process temperature

    max. 135°C (274°F)

  • Process pressure

    max. 10bar (145psi)

  • Temperature sensor

    Optional with integrated Pt100

  • Ex certification

    Simple operational equipment, no certification necessary.

  • Connection

    TOP68 connection head

  • Ingres protection

    IP 68

  • Additional certifications

    3A

Dokumenter / håndbøker / programvare

Tilbehør / reservedeler

Nr
Order code
Specification