Digital conductivity sensor
Condumax CLS16D

Digital conductivity sensor
Condumax CLS16D ©Endress+Hauser
Digital conductivity sensor
Condumax CLS16D ©Endress+Hauser
Digital conductivity sensor
Condumax CLS16D ©Endress+Hauser

The conductivity of a liquid can be measured using the conductive or toroidal measuring principles. This video shows what it is about and how these measuring principles work.

Memosens converts the measured value to a digital signal and transfers it to the transmitter without a contacting connection. This video shows how Memosens optimizes process control and maintenance.

Memosens conductive conductivity sensor for hygienic applications in pure and ultrapure water

Condumax CLS16D is a high-end conductivity sensor with certified hygienic design. It measures with highest precision providing you with reliable data for optimum process and product quality. Numerous certificates and various hygienic process connections ensure the perfect fit for your process. With Memosens technology, CLS16D combines maximum process and data integrity with simple operation. It resists corrosion and moisture, enables lab calibration, and facilitates predictive maintenance.

  • Fordeler

    • Highest precision and measuring accuracy

    • Hygienic design certified by EHEDG

    • Certified biocompatibility according to USP class VI and FDA compliance of all wetted materials

    • Sterilizable and autoclavable

    • Quality certificate stating the individual cell constant

    • Logging of sensor-specific data for easy traceability...

  • Bruksområde

    Condumax CLS16D measures conductivity in pure and ultrapure water applications for the monitoring and control of:

    • Ion exchangers

    • Reverse osmosis

    • Distillation

    • Electro-deionizing

    • Water for injection (WFI)

    With its international explosion protection...

Funksjoner og spesifikasjoner

Tilbake

Details

  • Measuring principle

    Conductive

  • Application

    Pure and ultrapure water, pharmaceutical industry

  • Characteristic

    Hygienic 2-electrode conductivity sensor

  • Measurement range

    k=0,1: 0,04 to 500 µS/cm

  • Measuring principle

    Conductivity cell

  • Design

    Hygienic 2-electrode conductivity sensor with coaxially arranged electrodes, electropolished stainless steel and FDA sealing material; sterilizable

  • Material

    Isolation: PEEK
    Electrode: polished stainless steel 1.4435
    Sealing: Isolast (FFKM) FDA conform

  • Dimension

    Electrode diameter: 17 mm (0.66 inch)
    Electrode length: 54 mm (2.13 inch)

  • Process temperature

    -5 to 120 °C (23 to 248 °F)
    For Sterilizsation: max. 150 °C at 5 bar
    (Max. 302 °F at 73 psi)

  • Process pressure

    13 bar at 20 °C (188 psi at 68 °F)
    9 bar at 120 °C (130 psi at 248 °F)

  • Temperature sensor

    NTC 30K

  • Connection

    Process connection: Varivent, Neumo, Clamp
    Cable: Memosens connector

  • Ingres protection

    IP68

  • Additional certifications

    Material certification 3.1
    EHEDG, pharma certificate of compliance
    FDA

Dokumenter / håndbøker / programvare

Tilbehør / reservedeler

Nr
Order code
Specification